Skip to main content

@ Pette Mini technical parameters

Measurement Results of @ Pette mini Single Channel Pipette (National Standard Verification Regulation/JJG 646-2006)

Vecverse @Pette Mini single-channel, variable volumeNational standard calibration regulations/JJG 646-2006
Order NO.Range of dispensingAccuracy assurance rangeVolumeVolume ErrorRepeatability ErrorVolume ErrorRepeatability Error
0120-00100.5μL-10μL0.5μL-10μL1μL±1.50%≤2.00%±12.0%≤6.0%
5μL±0.30%≤0.6%±8.0%≤4.0%
10μL±0.30%≤0.2%±8.0%≤4.0%
0120-01002μL-100μL5μL-100μL10μL±0.50%≤0.50%±8.0%≤4.0%
50μL±0.30%≤0.40%±3.0%≤1.5%
100μL±0.20%≤0.20%±2.0%≤1.0%
0120-02005μL-200μL20μL-200μL20μL±2.00%≤0.80%±4.0%≤2.0%
100μL±1.00%≤0.50%±2.0%≤1.0%
200μL±0.50%≤0.30%±1.5%≤1.0%
0120-03005μL-300μL30μL-300μL50μL±0.60%≤0.40%±3.0%≤1.5%
150μL±0.40%≤0.15%±2.0%≤1.0%
300μL±0.40%≤0.15%±1.5%≤1.0%
0120-100010μL-1000μL50μL-1000μL100μL±0.50%≤0.50%±2.0%≤1.0%
500μL±0.15%≤0.20%±1.0%≤0.5%
1000μL±0.15%≤0.20%±1.0%≤0.5%

The above test data are averages from multiple repeated tests conducted under specific environmental conditions at Shuye Laboratory.